谢炳熔, 廖俊强, 张年泉. 三明市2016年医用X射线机防护分析[J]. 职业卫生与应急救援, 2017, 35(6): 548-550. DOI: 10.16369/j.oher.issn.1007-1326.2017.06.018
引用本文: 谢炳熔, 廖俊强, 张年泉. 三明市2016年医用X射线机防护分析[J]. 职业卫生与应急救援, 2017, 35(6): 548-550. DOI: 10.16369/j.oher.issn.1007-1326.2017.06.018
XIE Bingrong, LIAO Junqiang, ZHANG Nianquan. Investigation on protective measures of medical X-ray machines in Sanming City, 2016[J]. Occupational Health and Emergency Rescue, 2017, 35(6): 548-550. DOI: 10.16369/j.oher.issn.1007-1326.2017.06.018
Citation: XIE Bingrong, LIAO Junqiang, ZHANG Nianquan. Investigation on protective measures of medical X-ray machines in Sanming City, 2016[J]. Occupational Health and Emergency Rescue, 2017, 35(6): 548-550. DOI: 10.16369/j.oher.issn.1007-1326.2017.06.018

三明市2016年医用X射线机防护分析

Investigation on protective measures of medical X-ray machines in Sanming City, 2016

  • 摘要:
    目的 了解三明市医用诊断X射线机放射防护情况与应用性能水平, 分析存在问题, 为做好放射防护工作提供科学依据, 进一步改善各级医疗单位放射防护质量。
    方法 按照《医用X射线诊断放射防护要求》《医用常规X射线设备影像质量控制检测规范》对三明市辖区内部分市属及县属公立医疗机构、乡镇公立医疗机构医用X射线机放射防护情况及应用性能进行调查。
    结果 三明市市属及县属医疗机构放射防护检测的医用诊断X射线设备共113台。放射防护机房周围剂量当量率合格率100%, 放射防护符合性评价"放射防护注意事项""灯箱处警示语句"和"工作状态指示灯与机房通道门有效联动"3个项目合格率低于50%;乡镇级医疗机构"工作状态指示类与机房通道门有效联动"和"个人防护用品配备"2个项目合格率显著低于市、县属医疗机构; 应用性能检测情况摄片机检测80台, 其中"光野与照射线周边的偏离""光野与照射野中心的偏离"和"管电压指示的偏离"3个项目合格率较低, 为60.00%~70.00%, 除"有用线束半值层"项目外, 乡镇级医疗机构其余项目合格率均低于市、县属医疗机构; 透视机及牙科X射线机应用性能也存在部分不合格项目。
    结论 各级医疗机构放射防护建设情况较好, 但相关设置及管理措施不够完善, 且应用性能合格率较低, 特别是乡镇级医疗机构。应加强宣贯, 加大经费投入, 整改不符合项, 并定期进行放射防护与应用性能检测。

     

/

返回文章
返回