王建宇, 张燕, 苏艺伟, 黄海波, 候狮峰, 张万峰, 刘移民. 13家医疗体检机构车载式诊断X射线机放射防护检测结果分析[J]. 职业卫生与应急救援, 2018, 36(4): 325-327. DOI: 10.16369/j.oher.issn.1007-1326.2018.04.012
引用本文: 王建宇, 张燕, 苏艺伟, 黄海波, 候狮峰, 张万峰, 刘移民. 13家医疗体检机构车载式诊断X射线机放射防护检测结果分析[J]. 职业卫生与应急救援, 2018, 36(4): 325-327. DOI: 10.16369/j.oher.issn.1007-1326.2018.04.012
WANG Jianyu, ZHANG Yan, SU Yiwei, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, LIU Yimin. Analysis of radiation protection effects of vehicle-mounted medical x-ray apparatus in 13 medical institutions[J]. Occupational Health and Emergency Rescue, 2018, 36(4): 325-327. DOI: 10.16369/j.oher.issn.1007-1326.2018.04.012
Citation: WANG Jianyu, ZHANG Yan, SU Yiwei, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, LIU Yimin. Analysis of radiation protection effects of vehicle-mounted medical x-ray apparatus in 13 medical institutions[J]. Occupational Health and Emergency Rescue, 2018, 36(4): 325-327. DOI: 10.16369/j.oher.issn.1007-1326.2018.04.012

13家医疗体检机构车载式诊断X射线机放射防护检测结果分析

Analysis of radiation protection effects of vehicle-mounted medical x-ray apparatus in 13 medical institutions

  • 摘要:
    目的 调查广州13家私营体检机构13台车载式诊断X射线机放射影像质量、防护性能参数和防护情况。
    方法 按照GBZ 264-2015、GBZ 130-2013和WS 76-2011等标准规范进行检测。
    结果 车载式X射线机房射线泄漏的点集中位于防护门缝、车体机房外左右两侧及车头处。13家单位的车载式诊断X射线机中, 有8家机构14处机房存在防护门门缝泄漏现象, 超过2.5 μSv/h, 最高达到6.64 μSv/h; 6家机构12处车体两侧超过2.5 μSv/h, 最高达到7.57 μSv/h, 3家机构车头处超过2.5 μSv/h, 最高达到4.7 μSv/h。其余机房外的周围剂量当量率均符合《医用X射线诊断放射防护要求》(GBZ 130-2013)的要求。防护性能方面, "管电压指示偏离" "输出量重复性""输出量线性"合格率均为92.3%, "光野与照射野四边的偏离"合格率为84.6%, 其余检测指标项目均为合格。
    结论 车载X射线机的防护门、车体两侧及车头处超标较为严重, 建议放射科医生每次作业时进行巡检。同时应加大自主监测和巡检的力度, 保障医疗放射人员的职业健康。

     

/

返回文章
返回