HUANG Haibo, WANG Jianyu, ZHANG Yan, SU Yiwei, ZHANG Wanfeng, HOU Shifeng, LIU Yimin. Effectiveness analysis of radiological protection measures against newly installed DR and gastrointestinal machine in a district hospital[J]. Occupational Health and Emergency Rescue, 2018, 36(5): 449-452. DOI: 10.16369/j.oher.issn.1007-1326.2018.05.021
Citation: HUANG Haibo, WANG Jianyu, ZHANG Yan, SU Yiwei, ZHANG Wanfeng, HOU Shifeng, LIU Yimin. Effectiveness analysis of radiological protection measures against newly installed DR and gastrointestinal machine in a district hospital[J]. Occupational Health and Emergency Rescue, 2018, 36(5): 449-452. DOI: 10.16369/j.oher.issn.1007-1326.2018.05.021

某区级医院新装DR、胃肠机放射防护控制效果评价

Effectiveness analysis of radiological protection measures against newly installed DR and gastrointestinal machine in a district hospital

  • 摘要:
    目的 分析与探讨广州市某区级公立医院新装DR、胃肠机、放射防护控制效果。
    方法 根据国家放射卫生有关法律、法规和标准, 通过现场放射卫生学调查和检测, 对该建设项目进行控制效果评价。
    结果 该医院的2间机房所在建筑物内的平面布局基本合理, 机房放射防护屏蔽参数及新建的辐射防护设施符合国家标准要求; 2间机房的周围剂量控制水平、年有效剂量约束值水平均符合标准要求; 2台设备性能参数符合标准要求; 2间机房的警示标志、通风情况、个人防护用品配备符合要求; 该项目制定的放射防护管理制度、建立的管理体系、职业健康管理情况符合标准要求, 但应细化安全联锁等相关操作制度; 未设置医患双通道, 胃肠机摆放不合理。
    结论 该院2个改建项目放射防护控制效果基本达到国家规定要求, 但后期应加强辐射防护监测。基层医疗机构应进一步完善放射防护管理, 以保证各项规章制度的落实。

     

/