WANG Jianyu, ZHANG Yan, SU Yiwei, ZHOU Liping, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, TANG Siqi, LIU Yimin. Analysis of monitoring results of X-ray imaging diagnostic equipment in 40 medical institutions in Guangzhou[J]. Occupational Health and Emergency Rescue, 2021, 39(6): 669-672. DOI: 10.16369/j.oher.issn.1007-1326.2021.06.013
Citation: WANG Jianyu, ZHANG Yan, SU Yiwei, ZHOU Liping, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, TANG Siqi, LIU Yimin. Analysis of monitoring results of X-ray imaging diagnostic equipment in 40 medical institutions in Guangzhou[J]. Occupational Health and Emergency Rescue, 2021, 39(6): 669-672. DOI: 10.16369/j.oher.issn.1007-1326.2021.06.013

广州市40家医疗机构X射线影像诊断设备监测结果及分析

Analysis of monitoring results of X-ray imaging diagnostic equipment in 40 medical institutions in Guangzhou

  • 摘要:
      目的  了解广州市各级医疗机构X射线影像诊断设备的性能和防护现状,保障医护放射从业人员的职业健康。
      方法  采用分层整群抽样的方法,调查广州市40家医疗机构的167台X射线影像诊断设备,依据相关的国家标准进行监测。
      结果  医疗机构的检测合格率为87.5%(35/40),X射线影像诊断设备质量控制监测合格率为88.0%(147/167)。骨密度仪、DSA、X射线摄影机合格率较高,CT机、X射线透视机合格率较低(P < 0.05)。143间机房外周围剂量当量率均未超过标准规定的限值,但是存在泄漏点,一、二、三级医疗机构监测点泄漏率差异有统计学意义(P < 0.05),其中以三级医疗机构的泄漏率最低。各级医疗机构防护区域重点为控制室防护门、机房防护门以及观察窗。
      结论  广州各级医疗机构X射线影像诊断设备整体监测结果良好,应根据不同级别医疗机构的放射诊疗管理方式,提出针对性意见,进一步提高自主管理能力,保障医护放射从业人员的职业健康。

     

/