王建宇, 张燕, 苏艺伟, 黄海波, 侯狮峰, 张万峰, 唐思琪, 刘移民. 广州市2020年25台数字减影血管造影设备机房放射防护分析[J]. 职业卫生与应急救援, 2022, 40(1): 70-73. DOI: 10.16369/j.oher.issn.1007-1326.2022.01.014
引用本文: 王建宇, 张燕, 苏艺伟, 黄海波, 侯狮峰, 张万峰, 唐思琪, 刘移民. 广州市2020年25台数字减影血管造影设备机房放射防护分析[J]. 职业卫生与应急救援, 2022, 40(1): 70-73. DOI: 10.16369/j.oher.issn.1007-1326.2022.01.014
WANG Jianyu, ZHANG Yan, SU Yiwei, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, TANG Siqi, LIU Yimin. Study on radiological protection against digital subtraction angiography equipment in Guangzhou in 2020[J]. Occupational Health and Emergency Rescue, 2022, 40(1): 70-73. DOI: 10.16369/j.oher.issn.1007-1326.2022.01.014
Citation: WANG Jianyu, ZHANG Yan, SU Yiwei, HUANG Haibo, HOU Shifeng, ZHANG Wanfeng, TANG Siqi, LIU Yimin. Study on radiological protection against digital subtraction angiography equipment in Guangzhou in 2020[J]. Occupational Health and Emergency Rescue, 2022, 40(1): 70-73. DOI: 10.16369/j.oher.issn.1007-1326.2022.01.014

广州市2020年25台数字减影血管造影设备机房放射防护分析

Study on radiological protection against digital subtraction angiography equipment in Guangzhou in 2020

  • 摘要:
      目的  了解广州市数字减影血管造影(digital subtraction angiography,DSA)设备机房的放射防护现状,为医疗机构保障放射介入工作人员的职业健康提供依据。
      方法  按照GBZ 130—2013《医用X射线诊断放射防护要求》等相关标准,对23家医疗机构的25台DSA设备进行防护检测,并对检测结果进行分析。
      结果  25台DSA设备透视防护区测试平面上空气比释动能率检测合格率为48.00%,其中第一术者位的检测结果为112.4(63.74,225.89)μGy/h,合格率为56.00%;第二术者位的检测结果为139.63(65.21,225.14)μGy/h,合格率为48.00%,第一术者位防护合格率高于第二术者位,差异有统计学意义(P < 0.05)。防护效果从高到低依次为头部、足部、下肢、胸部、腹部。工作场所外放射防护检测25间DSA机房,23间机房合格,合格率为92.00%。
      结论  DSA设备透视防护区的空气比释动能率较高,应加强防护,尤其应重视对第二术者位和第一术者位腹部的防护。应加强医疗机构放射防护知识培训,强调工作时佩戴放射防护个人防护用品的重要性,保障放射工作人员的职业健康

     

/

返回文章
返回