邓玮淳, 林潮鑫, 颜旭霖, 陈叙杰, 谢明珠. 胶体钯-石墨炉原子吸收法测定血铟方法研究[J]. 职业卫生与应急救援, 2023, 41(4): 512-515. DOI: 10.16369/j.oher.issn.1007-1326.2023.04.024
引用本文: 邓玮淳, 林潮鑫, 颜旭霖, 陈叙杰, 谢明珠. 胶体钯-石墨炉原子吸收法测定血铟方法研究[J]. 职业卫生与应急救援, 2023, 41(4): 512-515. DOI: 10.16369/j.oher.issn.1007-1326.2023.04.024
DENG Weichun, LIN Chaoxin, YAN Xulin, CHEN Xujie, XIE Mingzhu. Study on method for determination of blood indium with colloidal palladium graphite furnace atomic absorption spectrometry[J]. Occupational Health and Emergency Rescue, 2023, 41(4): 512-515. DOI: 10.16369/j.oher.issn.1007-1326.2023.04.024
Citation: DENG Weichun, LIN Chaoxin, YAN Xulin, CHEN Xujie, XIE Mingzhu. Study on method for determination of blood indium with colloidal palladium graphite furnace atomic absorption spectrometry[J]. Occupational Health and Emergency Rescue, 2023, 41(4): 512-515. DOI: 10.16369/j.oher.issn.1007-1326.2023.04.024

胶体钯-石墨炉原子吸收法测定血铟方法研究

Study on method for determination of blood indium with colloidal palladium graphite furnace atomic absorption spectrometry

  • 摘要:
      目的  使用胶体钯作为基体改进剂,优化石墨炉原子吸收法(graphite furnace atomic absorption spectrometry,GFAAS)测定血铟的方法。
      方法  采用纳米级胶体钯基体改进剂组合,用混合加入法测定血铟含量。
      结果  该法在0.55 ~ 40.00 μg/L的范围内有良好线性关系,相关系数0.999 1;方法最低检出限为0.55 μg/L;批内相对标准偏差为3.97% ~ 5.75%,批间相对标准偏差为5.20% ~ 6.05%;加标回收率为84.5% ~ 105.0%。测定血铟标准物质,结果在参考值范围内。
      结论  该法操作简便易行,胶体钯能有效消除基体干扰,准确度和精密度均达到生物材料检测的要求,在职业健康(生物材料)检测中应用此方法能达到批量、快速检测的目的。

     

/

返回文章