陈飚, 翟江龙, 李斌, 储呈晨, 姚杰, 陈晓文. 2019—2021年上海市77台医用CT稳定性检测结果分析[J]. 职业卫生与应急救援, 2024, 42(5): 624-627. DOI: 10.16369/j.oher.issn.1007-1326.2024.05.012
引用本文: 陈飚, 翟江龙, 李斌, 储呈晨, 姚杰, 陈晓文. 2019—2021年上海市77台医用CT稳定性检测结果分析[J]. 职业卫生与应急救援, 2024, 42(5): 624-627. DOI: 10.16369/j.oher.issn.1007-1326.2024.05.012
CHEN Biao, ZHAI Jianglong, LI Bin, CHU Chengchen, YAO Jie, CHEN Xiaowen. Results analysis of constancy tests of 77 diagnostic CT machines in Shanghai from 2019 to 2021[J]. Occupational Health and Emergency Rescue, 2024, 42(5): 624-627. DOI: 10.16369/j.oher.issn.1007-1326.2024.05.012
Citation: CHEN Biao, ZHAI Jianglong, LI Bin, CHU Chengchen, YAO Jie, CHEN Xiaowen. Results analysis of constancy tests of 77 diagnostic CT machines in Shanghai from 2019 to 2021[J]. Occupational Health and Emergency Rescue, 2024, 42(5): 624-627. DOI: 10.16369/j.oher.issn.1007-1326.2024.05.012

2019—2021年上海市77台医用CT稳定性检测结果分析

Results analysis of constancy tests of 77 diagnostic CT machines in Shanghai from 2019 to 2021

  • 摘要:
    目的 通过分析大型医用设备(CT)监控系统中CT稳定性检测结果,为管理部门进一步有效开展CT设备监管提供依据。
    方法 从监控系统中整理CT设备2019—2021年的检测结果,通过统计分析比较不同参数的变化情况。
    结果 本研究共纳入58家医疗机构77台CT各项指标的检测结果,所有结果均符合GB 17589—2011《X射线计算机断层摄影装置质量保证检测规范》规定的稳定性检测限值要求。各项参数中,CT值(水)、均匀性、重建层厚偏差、诊断床定位精度的检测值在不同年份间比较,差异均无统计学意义(P > 0.05),相对平稳;3年间噪声、高对比分辨力的检测值有一定波动(均P < 0.01),其中,2021年噪声检测值最低,2021年高对比分辨力检测值最高。
    结论 上海市CT设备各项性能指标稳定,监控系统有效促进了医用CT核心性能指标稳步提升,但其实用性仍有进一步完善的空间。

     

/