official WeChat

WANG Jun-sheng, ZHANG Yi, ZHAO Kun, LUO Jin-gan. Status of image quality control of conventional medical diagnostic X-ray equipment in Shenzhen[J]. Occupational Health and Emergency Rescue, 2013, 31(3): 129-131.
Citation: WANG Jun-sheng, ZHANG Yi, ZHAO Kun, LUO Jin-gan. Status of image quality control of conventional medical diagnostic X-ray equipment in Shenzhen[J]. Occupational Health and Emergency Rescue, 2013, 31(3): 129-131.

Status of image quality control of conventional medical diagnostic X-ray equipment in Shenzhen

More Information
  • Received Date: April 10, 2013
  • Revised Date: May 15, 2013
  • [Objective] To investigate the status of the image quality control of conventional medical diagnostic X-ray equipment in Shenzhen. [Methods] Multi-functional X-ray detector and X-ray machine test plate were used to test the main imaging quality control indicators of X-ray machine. The data were compared with the relevant standards in China. [Results] Totally there were 364 medical diagnostic X-ray machines in 167 medical institutions. The total qualified rate was 88.74%. However, there was a great difference of qualified rate in different sub-groups of conventional medical diagnostic X-ray equipment. The lowest was common X-ray photographs, in which the bigger edges deviation of lighting field to radiation field has serious problem. [Conclusion] We should strengthen the testing and management of imaging quality control of medical X-ray diagnostic equipment to protect the safety and health of patients.
  • [1]
    郑钧正.电离辐射医学应用的防护与安全[M].北京:原子能出版社,2009:15.
    [2]
    中华人民共和国卫生部. WS 76-2011医用常规X射线诊断设备影像质量控制检测规范[S]. 北京:中国标准出版社,2011.
    [3]
    UN SCEAR. Sources and effects of ionizing radiation[R]. New York:UN,2000.
    [4]
    郭常义,高林峰,张力,等. 上海市医用诊断X射线机质量状况调查和评价[J]. 中国辐射卫生,2008,17(1):61-63.
    [5]
    ICRP. ICRP Publication 105,Radiological protection inmedicine[M].Oxford:Pergamon Press,2008.
    [6]
    张静波,邹蓉珠,张林,等. 广州市医用诊断X射线机影像质量状况调查和分析[J]. 中国辐射卫生,2009,18(3):328-330.
    [7]
    格日勒满达呼,哈日巴拉,许潇,等. 内蒙古中西部地区医用X射线诊断机质量控制检测结果分析[J]. 中国辐射卫生,2011,20(4):443-444.
    [8]
    赵雅斐,张鹏,郭伟,等. 郑州市医用诊断X射线机质量控制检测与评价[J]. 中国辐射卫生,2008,17(3):321-322.
    [9]
    王全锋,李举跃,陈小霞. 濮阳市医用常规X射线诊断设备影像质量控制检测结果及分析[J]. 中国辐射卫生,2012,21(2):196.
    [10]
    杜世民. 三门峡市医用诊断X射线机影像质量控制现状调查[J]. 河南预防医学杂志,2012,23(3):214-215.

Catalog

    Article views (249) PDF downloads (39) Cited by()

    /